|
Radiation Safety Office Henry Ford Health System
3. Fluoroscopy System Description and Operation
|
|
|
Chapter 3: Fluoroscopy System Description and Operation Figure 1-2: Fluoroscopy System Components Courtesy of Scott Sorenson, 2000
The X-ray generator controls the quantity (number) and quality (spectrum of energy) of the X-rays produced. There are three basic controls to the generator: kVp - voltage applied accross the X-ray tube. mA -Current across the X-ray tube. Time - starting and stopping the exposure. From a safety perspective, the beam- on- time is the most important control. Beam-on-Time (Foot Pedal/Switch) Radiation exposure during fluoroscopy is directly proportional to the length of time the unit is activated by the foot pedal or switch. Unlike regular X-ray units, fluoroscopic units do not have an automatic timer to terminate the exposure after it is activated. Instead, depression of the switch determines the length of the exposure which ceases only after the switch is released. Thus, cognizance of beam-on-time is extremely important to fluoroscopy safety. X-ray tube Current (mA) The X-ray tube current (mA setting) essentially controls the quantity of X-rays produced per unit time. If you double the mA, you double the patient and operator exposure. The higher current mode (High Level Control or "boost" mode) dramatically increases the number of X-rays produced and thereby improves image quality. The boost mode is activated in some systems when the foot switch is firmly depressed (a light push activates the normal, low dose, mode). An audible alarm will indicate when the high level mode is being used. HFH policy limits the rate for the boost mode to 20 R/minute (compared with maximum of 10 R/minute for the normal, Automatic Brigtness Control (ABC) mode and 5 R/minute for the manual mode). Both the patient and fluoroscopy operator dose are proportional to the total amount of current used by the machine. Thus, a light touch on the pedal (educated use of the boost mode) will minimize both patient and worker doses.
During normal mode fluoroscopy, the average patient Entrance Skin
Exposure (ESE) is approximately 2 R/min. The level
of radiation exposure falls off exponentially with increasing tissue depth due to attenuation and
inverse-square effects. Only approximately 1% of the original radiation beam reaches the
I-I for image generation. Fluoroscopy Timer Fluoroscopy Monitoring The Radiation Safety Committee requires that the total elapsed fluoroscopy time be recorded for every procedure that exceeds 10 minutes of fluoroscopy. Certain areas, primarily in the Henry Ford Hospital, are also required to record all fluoroscopy times. Sheets used to record fluoroscopy times are available from the Radiation Safety Office. Imaging Modes
Automatic Brightness Control (ABC) Figure 3-2: Automatic Brightness Control System Courtesy of Phil Rauch, 2000
Obese patients drive the X-ray machine output up considerably. Thus, obese patients have the greatest risk of skin injury. Magnification Modes Figure 3-3: Field of View Courtesy of Scott Sorenson, 2000
Normal Magnification Mode The Food and Drug Administration (FDA) regulates the construction of all fluoroscopy systems. For routine fluoroscopy applications, the FDA limits the maximum ESE to 10 R/min. The use of higher radiation rates ("High Level Control" or "boost" modes) are useful in situations requiring high video image resolution. ESE of up to 20 R/min is permitted for short duration. Special operator reminders, such as audible alarms, are activated during "boost" modes. Figure 3-4 illustrates the effect of changing Field-Of-View, or magnification modes, on skin entrance exposure (ESE) for a typical fluoroscopy system: Figure 3-4: Image Intensifier Input Exposure
Courtesy of Phil Rauch, 2000 Cineangiography
Cineangiography (cine) originally involved exposing cinematic film to the
I-I output, providing a permanent record of the imaged sequence. Technological advances in film-less imaging have
eliminated the film for most systems. The cine mode extracts several
separate diagnostic quality images per minute. The amount of information
collected for each of these images is
essentially equivalent to a normal flat plane X-ray image. The X-ray
machine output required to produce a cine movie is much higher than the level needed for
normal fluoroscopy.
Consequently, dose rates during cine image collection are usually 10 to 20 times higher than normal
fluoroscopy. For this reason,
careful use of cineangiography is required. Figure 3-5: X-ray tube and Collimator System Courtesy of Phil Rauch, 2000
Figure 3-6: Benefits of Collimation Courtesy of Scott Sorenson, 2000
Prudent use of collimators can also improve image quality by blocking-out "bright areas," such as lung or other low density regions, allowing better resolution of other tissues.
Benefits from using collimation
Figure 3-7: Collimation Technique Courtesy of Scott Sorenson, 2000
Last Image Hold Beam Quality (kVp) The tube voltage (kV) controls the maximum energy of electrons produced by the X-ray tube and the maximum energy of the resultant X-ray spectrum (kilovolt-peak energy or kVp). Use of high kVp techniques can reduce patient dose but contrast between differing tissues is also reduced. Experienced operators can optimize the choice between image quality and patient dose through careful adjustment. The ABC on most X-ray systems can change the kVp and mA used to optimize imaging. Image Display Monitor The image quality available to the operator is dependent on proper adjustment of the image monitor. These image monitor settings are adjusted during service and annual testing. Operator adjustment of the brightness and contrast controls can degrade image sharpness, contrast and distortion and should be done if problems occur during use. If this happens, service should be contacted to correct the problems. In addition, the eye's ability to discern detail on the image monitor is improved under low light conditions. The need for good lighting for surgical needs must be balanced with imaging considerations. Therefore, the monitor settings should be calibrated for the area in which it is being used. System Quality Control Checks Quality Control (QC) checks are extremely important to ensure proper performance of the equipment. Daily testing is required by JCAHO standards and should be utilized to track system performance and optimize display monitor settings. A simple test tool (Fig 3-8) can be used to monitor system performance. This tool should be used to evaluate the X-ray system each day prior to use and any discrepancies corrected immediately. Figure 3-8: Quality Control Checks Courtesy of Phil Rauch, 2000
Operator Exposure Profile Figure 3-9: Collimation Technique
Courtesy of Scott Sorenson, 2000
It is important to note that the patient does not uniformly emit this scattered radiation since some of the scattered radiation is absorbed or reduced in intensity by passing through the patient. The intensity of scatter decreases with increasing distance, due to inverse square law effects (see Chapter 1). Consequently, scatter radiation is highest near its source (i.e. the X-ray beam entry point on the patient). Because radiation scattered in the forward direction (into the patient) is subject the most tissue attenuation, radiation levels are significantly lower on the I-I side than the X-ray tube side. Highest scatter radiation levels are often where the operator stands. Radiation levels increase with decreasing distance from the point of X-ray entry (Figure 3-10). In general, an operator positioned 3 feet from the X-ray beam entrance area will receive 0.1% of the patient’s ESE. Staff members positioned further away receive much less exposure due to inverse square law effects. In almost all cases, the tableside operator will receive the highest occupational radiation exposure during the fluoroscopic procedure. Tableside fluoroscopy receive among the highest occupational radiation exposures within the health system. Figure 3-10: Radiation Level vs. Entry Point Courtesy of Scott Sorenson, 2000
Radiation levels are highest beneath the table (when the X-ray tube is below the patient) because the patient provides an effective beam stop (Figure 3-11). Highest levels are directed at the operator's waist (See bar chart on figure). Figure 3-11: Patient Shielding
Courtesy of Scott Sorenson, 2000
Figure 3-12: Tilt Exposure Profile
Courtesy of Scott Sorenson, 2000 Courtesy of Scott Sorenson, 2000 Effect of rotating X-ray system. Images taken with the I-I away (Figure 3-12) result in higher radiation exposure to the operator's eyes compared to images with the I-I towards the operator (Figure 3-13). Figure 3-13: Tilt Exposure Profile Courtesy of Scott Sorenson, 2000
A basic principle in Radiology is that you collect dose along with image information. The clearest, least jittery, images produce the highest doses. Thus, it is very important for clinicians to judiciously use appropriate judgment when increasing the X-ray beam output and to learn to work with most amount of (necessary) imaging imperfections as possible which still allow the needed clinical outcome. |
|
Copyright © 2001
Radiation Safety Office at Henry Ford Health System
|